Service and maintenance
6.3 Installing and removing electronic components
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
93
2.
Rotate the 1.5 mm hexagon socket-head bolt in the cover
②
in the clockwise direction as
far as possible.
WARNING
Tightening the locking screw
According to the explosion protection regulations, all housing covers which have to be
unscrewed to open the housing must each be secured against unintentional opening by
a locking screw
②
.
Before opening, loosen this hexagon socket-head bolt
②
.
Following closing, tighten this hexagon socket-head bolt
②
.
3.
The system parameters must be downloaded when the device is switched on again. The
procedure is described in the "Maxum System Manager" operating software manual.