Service and maintenance
6.3 Installing and removing electronic components
MicroSAM
88
Operating Instructions, 06/2012, C79000-G5376-C560-07
6.3.1
Removal of electronic modules
Replacing the backup battery
The gas chromatograph has a backup battery for the built-in clock.
WARNING
Explosion hazard
Observe the regulations for explosion protection if the chromatograph is installed in a
hazardous area. Replacement of this backup battery is only permissible in the hazardous
area if you have a fire permission certificate. This applies whether the gas chromatograph
is switched on or off.
Procedure
1.
Remove the connector
②
on the battery module.
2.
Loosen the screw T30
①
.
3.
Remove the backup battery
③
.
4.
Install the new backup battery in the reverse order.