Technical data
MicroSAM
106
Operating Instructions, 06/2012, C79000-G5376-C560-07
Electromagnetic compatibility
EMC immunity
In accordance with IEC 613261-1:2005, NAMUR NE 21
Conducted interference on the AC voltage supply lines:
IEC 61000-4-4
2 kV
IEC 61000-4-5
Asymmetric/symmetric 2 kV/1 kV
IEC 61000-4-6
10 V
IEC 61000-4-11
passed
Conducted interference on signal lines:
IEC 61000-4-4
1 kV
IEC 61000-4-5
Asymmetric/symmetric 1 kV/0.5 kV
IEC 61000-4-6
10 V
Immunity to discharges of static electricity:
IEC 61000-4-2
Air/contact 8 kV
Immunity to fields:
IEC 61000-4-3
10 V/m
Emitted Interference:
CISPR 11/EN 55011
Class B (living accommodation)
KC marking for use in Korea only
A급 기기
이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을
주의하시기 바라며 가정 외의 지역에서 사용하는 것을 목적으로 합니다.
Safety
Electrical safety
IEC 61010 / DIN VDE 0411
Protection against explosion
ATEX and IEC Ex: II 2 G Ex d IIC T4
Class I, Zone 1, Group IIB + H2 T4
Class I, Div 1, Groups B, C, D T4
Factory Sealed
Содержание MicroSAM
Страница 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Страница 2: ......
Страница 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 72: ...Mounting 4 4 Mounting and connecting MicroSAM 70 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 78: ...Commissioning and operation 5 5 Shutting down MicroSAM 76 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...