Service and maintenance
6.4 Conversion of MicroSAM basic device AS09
MicroSAM
98
Operating Instructions, 06/2012, C79000-G5376-C560-07
6.4.3
Returning for use of analysis modules of type D
Procedure
1.
Terminate measuring mode of the device.
2.
Deactivate the detectors using the operating software.
3.
Save the configuration in the flash memory.
4.
Switch off the device.
5.
Shut off the carrier gas and sample gas supplies.
6.
Open the housing.
7.
Deinstall the analysis module type A.
8.
Install the analysis module type D.
9.
Remove the RSP module.