Mounting
4.4 Mounting and connecting
MicroSAM
68
Operating Instructions, 06/2012, C79000-G5376-C560-07
3.
Then carefully lever the insert out of the chromatograph as shown in the following picture.
A 25-contact Sub-D socket becomes visible.
4.
Connect the Ethernet connection of the HMI PC to the corresponding pins of the
chromatograph according to the following table. If the display board is removed, the
Ethernet connection via the standard cable is interrupted.
Comment
Pin No. Description
Mounting with hub
1)
Mounting without hub
2)
22
Ethernet 10BaseT TX+
Ethernet cable:
Orange/white
Ethernet cable:
Green/white
23
Ethernet 10BaseT TX-
Ethernet cable:
Orange
Ethernet cable:
Green
24
Ethernet 10BaseT RX+
Ethernet cable: Green/white Ethernet cable:
Orange/white
25
Ethernet 10BaseT RX-
Ethernet cable:
Green
Ethernet cable:
Orange
1)
Standard EIA/TIA-T568A assignment
2)
Standard EIA/TIA-T568B assignment
Following completion of on-site operation
1.
Screw in the securing ring as far as possible.
Содержание MicroSAM
Страница 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Страница 2: ......
Страница 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 72: ...Mounting 4 4 Mounting and connecting MicroSAM 70 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 78: ...Commissioning and operation 5 5 Shutting down MicroSAM 76 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Страница 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...