8.1.6
Totalizer
Totalizer function
The device has one totalizer function that can be used to totalize the mass flow process value.
Default unit of the totalizer can be changed, see Modbus register 8320.
The totalizer may be paused, resumed or reset:
• Pause (Modbus address 2613): the totalizer holds the last value before the failure occurred
• Resume (Modbus address 2614): the totalizer continues counting the actual measured value
• Reset (Modbus address 2612): the totalizer continues counting based on the last input value
(for example mass flow) before the failure occurred.
Note
The totalizer is reset on power failure.