Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2442
Float / 4
Density Correction
Factor
Set density compensation
value (gain) in order to make
a density correction (scale
factor).
To increase the displayed
density value with +0.5 %,
set the density factor to
1.005.
The displayed density value
will now be 0.5 % higher
than before
1
-1.999 to
+1.999
Read /
Write
2444
Float / 4
Density Correction
Offset
Set density compensation
value (offset) in order to
make an offset on the meas‐
ured density.
To make the flowmeter show
+ 2 kg/m
3
, change the densi‐
ty offset to 2.000 kg/m
3
in the
'Sensor' menu
0 [kg/m
3
]
-1 400 to +1
400
Read /
Write
Table C-8
Totalizer
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default
value
[unit]
Value range
Access
level
2609
unsigned / 2 Totalizer State
Totalizer state
• 0 = paused
• 1= running
1
0 to 1
Read only
2610
float / 4
Totalizer Value
The totalized MASS value in
kg
0 [kg]
Min. -1.70E+38
Max. 1.70E+38
Read Only
3018
Unsigned / 4 Totalizer fixed
point part
The totalized MASS value in
kg. Most significant word of
the totalizer (MSW).
The format of the totalizer is
a TotalType.
The TotalType represent a
fixed point value in the MSW
32 bit and a fractional part in
the LSW 32 bit.
Example: 2.03 would be rep‐
resented as fixed point part =
2 and fractional part =
30000000
0 [kg]
Min
-2147483648
Max
2247483647
Read Only
Modbus communication FCT010
C.4 Modbus holding registers tables
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
165