Functions
7.9 Analog Output Trim
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
97
3.
To enable numeric entry, press the <Right Arrow>. Type a numeric value corresponding
to a 4mA input signal.
4.
To store the data press <ENTER>. This moves the cursor to [20 mA].
5.
To enable numeric entry, press the <Right Arrow>. Type the numeric value
corresponding to a 20mA input signal.
6.
To store the data, press <ENTER>.
7.9
Analog Output Trim
Introduction
Analog Out Trim function allows you to fine-tune the flow meter’s analog voltage and current
outputs using a multi-meter connected to the output under test. In addition, you can use a
frequency counter to fine-tune the flow meter’s pulse rate output.
Note
The current, voltage, and Pgen trimming will be limited by the 12-bit resolution of the flow
meter’s D/A Convertor (DAC).
1.
From the [Meter Facilities] menu, scroll to the [Analog OutTrim] menu.
2.
Press the <Right Arrow> to access the option list.
0HWHU)DFLOLWLHV
6LHPHQV
'XDO3DWK>@3DWK
$QDORJ2XWSXW7ULP
3UHIHUUHG8QLWV
(QJOLVK
7DEOH6HWXSV
/RJJHU&RQWURO
0HPRU\&RQWURO
$QDORJ2XW7ULP
57'&DOLEUDWH
&ORFN6HW
566HWXS
1RQH>@
%DFNOLJKW
2Q
6\VWHP,QIR
Table 7- 8
Analog Out Trim Menu Structure
Trim Io1/Io2
Operate
Trim@4mA Indicated mA=x.xx
Analog Out Trim
Trim Vo1/Vo2
Operate
Trim@2V Indicated V=x.xx
Trim Pgen1 / Pgen2
Operate
Trim@1 Hz Indicated Hz=xxxx
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......