Functions
7.3 Logger Control
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
85
4.
To transmit Logger contents to external device via the serial port press <ENTER>.
5.
To stop printout press <Left Arrow>.
Circular Memory
In its default mode, the Logger collects data until its memory becomes full. At that time the
flow meter suspends datalogging and cannot resume until the Logger memory is cleared
(see Clear Logger command). Circular Memory allows the Logger to "‘write over" its oldest
records when memory reaches full capacity. If you enable [Circular Memory], you are
assured of always collecting the most recent data. But also remember that you will lose the
oldest Logger reports and that further invoking of [Circular Memory] deletes the current
contents of the Logger.
To setup and enable Circular Memory:
1.
The Logger Mode menu must have the [Memory] menu cell selected.
2.
Logger items must be selected (e.g., Site ID, Date, Time, etc.).
3.
All active channels/paths in the Channel Setup menu must be disabled. To disable active
channels, select the [Channel Enable] menu cell and then [No].
4.
In the Logger Control menu, select [Circular Memory].
5.
Press <Right Arrow> to access the [Circular Memory] option list.
6.
Move the cursor to [Yes] by pressing <Up/Down Arrow>.
7.
To store selection press <ENTER>.
8.
Lastly, re-enable the channels/paths that you disabled earlier to begin logging.
Est LogTime Left
Est LogTime Left is a "view-only" menu cell that shows an estimate of the amount of Logger
time remaining in hours and minutes. This menu cell becomes active after you enable
datalogging. Selecting [Circular Memory] and/or event-based datalogging (see Logger
Setup), blanks the [Est LogTime Left] field and is based on the log interval and data
selections made in the Logger Setup.
Clear Logger
If you use the Logger in its default mode, eventually you will use all the memory available for
Logger storage. When this occurs, you will not be able to log more data until you free up the
memory. The [Clear Logger] command erases ALL stored Logger data. Therefore, you
should evaluate the currently stored data, and print any valuable information before using
this command.
Note
Saved Sites also consume Logger RAM.
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......