10
Troubleshooting
10.1
Troubleshooting
The following is list of troubleshooting tips and messages that you may encounter. They
include explanations and, in some cases, a recommended action. If a problem seems
unsolvable, contact your local Siemens office or regional Ultrasonic Flow Representative for
expert help at: http://www.automation.siemens.com/partner
(
http://www.automation.siemens.com/partner
).
Table 10- 1 Troubleshooting Tips
Error or Message
Probable Cause
Solution
Memory Full!
Response to an attempt to save site data,
when data memory is full.
Delete an obsolete site or clear Datalogger
memory to make room for the new data.
Memory Corrupted!
Memory read error occurred while
accessing the active site data.
Refer to F4 reset procedure in the Operation
Instructions manual.
Chan Not Setup
Response to an attempt to invoke an
operation that requires a channel to be
enabled.
Enable the channel [Channel Setup - Channel
Enable - Yes]. Note that a channel cannot be
enabled until an "Install" operation is completed.
Clr Active Memory?
Response to pressing and holding the F4
key during power-up.
Use the F4 key function to restore operation if a
severe event (e.g., a violent power surge) disrupts
system operation.
Clr Saved Data?
[Clr Saved Data?] only appears after
pressing the <Down Arrow> in response to
[Clr Active Memory?].
Answering Yes to [Clr Saved Data?] will erase ALL
saved data. To invoke in RS-232 serial mode, type
@@@ and then press <ENTER> key.
<EOT>
Response to a request to output Datalogger
data to the printer or the Graphics screen
when no Datalogger data exists or at the
end of a transmitted file..
Set up the Datalogger.
No Sites - Press
<ENTER>
Response while trying to recall/delete a site
setup when no sites are stored.
Create a site.
Security
Response upon changing previously
entered data when security switch is in
[Disable] position or security code has been
entered.
Change switch position to [Enable].
Enter previously set security code.
RTC Error
Component level problem.
Meter requires service. Request RMA.
- - -F- - - Fault Alarm
Loss of signal strength (ALC)
Change of Rx signal location (Beam
Blowing)
Recouple sensors with fresh couplant.
Install sensors in Direct mount mode
Note: If problem persists call Tech support.
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
123
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......