Functions
7.8 Analog Input Setup
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
95
For example, when using the analog input Aux function, assigned numeric variables for
temperature monitoring might be spanned as follows: 4mA=1 (0°C / 32°F) and 20mA=100
(100°C / 212°F). Temperature data sent to the Aux 1 input port will feed through the flow
meter and report the scaling data to the Datalogger.
Although the flow meter receives these signals, the standard operating system will not use
them as available in any of its operations. The analog input only acts as a convenient "feed-
through" to add data to Datalogger reports.
Note
There will be no numeric units shown next to the output data. Be aware of the appropriate
units when entering scaling data and when viewing the resultant output on the Datalogger
report.
Note
Refer to the Installation Drawings or I/O Module markings for the locations of these inputs
and wiring procedures.
Input
Off
Aux
4 mA
Numeric entry
Iin1
20 mA
Numeric entry
Iin2
Same as Iin1
Input
Off
Aux
0 Volts
Numeric entry
Vin1
10 Volts
Numeric entry
I/O Data Control
Analog Inp Setup
Vin2
Same as Vin1
The flow meter recognizes the first analog input variable that is assigned to any given
parameter and ignores any subsequent input with the same assignment.
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......