Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
130
Operating Instructions, 03/2012, A5E02951522-02
To activate the Test Facilities Graph Screen:
1.
In the main menu, scroll to the [Diagnostic Data] menu and select [Test Facilities].
2.
Scroll down to [Graph], press the <Right Arrow> and highlight [Yes]. Press <ENTER> to
select.
3.
The Test Facilities Graphic Screen will appear on the meter display as shown below.
)
96
>@
'
7Q
G7
6
61
0LQ'DPS
①
Damping Factors
④
Min Damping Factor (Hot Key 1)
②
⑤
Digitized Receive Signal
TN Marker
③
Crossover Marker
⑥
High Baseline Noise
Figure 10-2 Setting Digital Damping Factor
Setting the Digital Damping Factor to a value HIGHER than the default value of 4 may be
necessary in cases where the signal-to-noise ratio (SN) is found to be unacceptably low
(<15:1), but only if the noise is determined to be asynchronous (i.e., not associated with the
transmit or flow meter timing circuitry) as shown in the signal example above, where the
baseline noise has a higher frequency than the true liquid signal.
The following application conditions may require a higher Digital Damping Factor:
●
Close proximity to pressure control valves which may generate in-band acoustic noise
●
High un-dissolved gas solids content in liquid.
●
High electronic noise from variable frequency drives or other external equipment.
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......