Initiator
Text
Classifica‐
tion
Description
Possible causes/remedy
CDR/CDH
EL. too hot
M
aint.
Electronics too hot.
•
Allow device to cool down.
•
Ambient temperature too high?
Air purge low
M
aint.
Volume flow below set limit.
Check purge air supply.
Filter watch
M
aint.
Flow monitor.
p no signal
M
aint.
No signal from pressure sensor.
p out of range
M
aint.
Sample gas pressure < 500 or > 1200 hPa
(mbar).
---
t air no signal
M
aint.
Broken sensor
Please contact SICK Customer Service.
[t] no signal
M
aint.
EEPROM
defect
M
aint.
EEPROM defective.
Heat no signal
M
aint.
Heater fault.
Heater < 1.5 A
M
aint.
Visor signals outside the valid range.
Heater defect
M
aint.
All 4Q signals below threshold parameter.
Heating too
low
M
aint.
Lamp does not go on.
No com.
M
aint.
Communication error to optic head and/or
reflector.
Communication error to optic head and/or
reflector.
System
Systemstart
X
tended
This message is entered during each system
start.
Shows when last system start was made.
Zero adjust
X
tended
The start of an adjustment is recorded in the
logbook.
Shows when last adjustment was made.
Boxmeasuring
X
tended
The start of filter box measurement is
recorded in the logbook.
Shows when last filter box measurement was
made.
Reflector
search
X
tended
Search for reflector failed.
•
Check alignment,
(option) - automatic optical alignment
check", page 69
•
Reflector contaminated or defective.
•
Light intensity weakness too strong on
measuring path.
P
Substitute
value
M
aint.
Calculation is made with a substitute value
because of a pressure measurement error.
T
M
aint.
H
M
aint.
Table 13: Error table: Maintenance request / Xtended
TROUBLESHOOTING
10
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
91