Measuring
2
Zero point
3
Check
4
Data recorder
5
Swivel element
6
Zero point reflector
7
Zero point
8
Check point (70% fsc)
Table 3: Measuring principle of check cycle
Output of measured values during the check cycle
Last valid measured value
Signal output during the check cycle (optional digital output or
OPC interface)
Not_measuring
Output of determined zero and reference values on analog
outputs
•
Directly after the check cycle
•
On request via a digital input (option)
Signal output during the output
Output_control values (optional digital output or OPC inter‐
face)
Output sequence:
1
Zero values for 90 s
2
Reference values for 90 s
Display of zero and reference values of the last check cycle as
well as QAL3 values in SOPAS ET
Menu: Diagnosis/check values
Display when check of NO cell failed
•
All interfaces: Results of NO cell.
•
All interfaces: "0" is output instead of zero and reference
values.
•
Analog output: "Live Zero".
•
Results of zero and reference measurement are irrele‐
vant.
PRODUCT DESCRIPTION
3
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
19