7.5
Aligning the measuring probe in flow direction
Flow direction aligned at the factory
NOTE
The fitting angle of the probe is already set before delivery when the gas flow direction
has been defined during project planning.
A sticker marks the setting.
1
2
3
Figure 23: Flow direction marking
1
Device flange
2
Flow direction sticker
3
Mounting ring screws (4 pcs.)
Measuring probe, aligning
Rotate the device flange to align the probe.
1
Loosen the 4 screws on the mounting ring (see Figure).
2
Rotate the device flange to the following position:
–
The measuring gap must point in flow direction.
–
The device flange must be positioned so that the SR-unit can be fitted in a
vertical position.
3
Tighten the screws on the mounting ring again to fasten the device flange in this
position.
7.6
Connecting protective gas to the SR-unit
The protective gas connection is on the intermediate enclosure.
7
COMMISSIONING
54
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
8015558/YYJ8/V1.7/2016-07 | SICK