1
2
3
1
2
Figure 37: Loosen the screws and replace the activated charcoal filter bag
1
Screw (1 of 6)
2
Activated charcoal bag
3
Spring
1
Take the SR-unit off,
see "Swiveling out and removing the SR-unit", page 75
.
2
Place the SR-unit in a clean, dry place with little dust.
3
Loosen six screws. (see Figure)
4
Take the cover off to the rear.
5
Replace the used activated charcoal bag with a new one.
6
Replace the cover and screw tight.
7
Refit the SR-unit.
MAINTENANCE
9
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
79
Subject to change without notice