7.8
Installing the measuring probe in the gas duct
CAUTION
Damage when device is dropped
Its weight and size make the SR-unit with installed probe hard to handle. This increases
the risk that the measuring device is damaged when dropped.
b
Install the SR-unit and the probe separately.
1
Insert the measuring probe with purge air fixture or flange attachment (without SR-
unit) in the “flange with tube” on the duct side.
2
Screw the measuring probe with purge air fixture or flange attachment tight to the
“flange with tube” (seal and 4 screws).
1 2
3 4
5
6
7
Figure 26: Fitting the probe on the measuring channel
1
Screw
2
Washer
3
Spring sheet
4
Nut
5
Purge air fixture
6
Seal
7
Flange with tube