Check cycle 1.1.4
/Menü/Diag/Check cycle
Measuring
Back
1 / 3
04/29 01:57
SO2:
Zero -5.0 0.5%
Span 699.8 -0.0%
1
3
4
5
6
7
2
Figure 32: Description of the check cycle display
1
Duct (in the example: Duct 1 of 3, according to. configuration)
2
Date (MM/DD) and time of the last check cycle
3
Gas component
4
Zero adjust result: Absolute value (in unit configured)
5
Deviation ín percentage of FS
6
Span adjust result (70%): Absolute value (in unit configured)
7
Deviation ín percentage of FS
1
1
FS = full scale value: Scale limit value of the assigned analog output
8.3.4.3
Alignment check (option) - automatic optical alignment check
This menu serves to view the values of automatic optical alignment.
NOTICE
b
Perform this check only when the SR-unit is at operating temperature (in operation
for at least 30 minutes).
b
Automatic mirror adjustment - do not perform manual adjustments.
NOTE
Further information on optical alignment,
see "Alignment adjust - manual optical align‐
Toggle between "Deviation" and "Performed steps of tracking mirror"
b
Press the
Down arrow button
.
b
Exit the menu item with the
Back button
.
OPERATION
8
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
69