3.6
Flow recalibration
To recalibrate the FLOWSIC600 at a flow test facility, apply the procedure as described in
Section
by using the "Meter calibration" wizard in
MEPAFLOW600 CBM.
Check the remaining battery capacity and replace the battery if necessary (
).
Set the current gas conditions on page 1 of 4 (
Perform the functional check (
).
Go to page 2 of 4 of the "Meter calibration" wizard.
Select "None" as adjustment method. Now the flow meter does not use the adjust fac-
tors displayed on the right hand side. The factors are still displayed and not deleted so
that they can be reactivated (see Note below).
Figure 42
"Meter calibration" wizard page 2 of 4, calibration factors "as found" (non flow calibrated flow meter)
Now proceed with
of this Chapter.
NOTICE:
Special handling procedures for recalibration defined in standards,
recommendations, national regulations and or type approvals must be
considered.
If it becomes apparent after the flow tests, that the adjust factors do not have
to be changed, do not write the parameter to the meter, but reactivate the
previous adjust factors on page 2 of 4 in the wizard, by selecting the previously
used adjustment method (other than "None").
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...