FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Commissioning
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Table 24
Diagnostics Comparison limits
Monitored difference
between reference
values and current
values
Configurable
difference limit
Default
value
Notes
Default
activation
state
1
Profile factor
Profile Factor
change
10%
A change of the profile factor value may be caused by contamination,
blockage or deposits in the line that change the symmetry of the flow
profile.
We recommend to use the default value.
Off
Symmetry
Symmetry
change
10%
A change of the symmetry value may be caused by contamination,
blockage or deposits in the line that change the symmetry of the flow
profile.
We recommend to use the default value.
SOS differences
between paths
SOS difference
change
1%
The SOS deviation indicates whether or not a path is measuring the
correct transit time.
We recommend to use the default value.
Turbulence
Turbulence
change
50%
A change of the turbulence value may be caused by contamination,
blockage or deposits in the line that change the symmetry of the flow
profile.
We recommend to use the default value.
SNR (Signal-to-noise
ratio)
SNR change
20dB
Interfering noise caused by fittings in the pipeline, valves that are not
fully open, sources of noise near the measuring location, or defective
ultrasonic transducers may affect the signal-to-noise-ratio.
We recommend to use the default value.
AGC (Signal
amplification)
AGC change
10dB
If the AGCs of a path deviate more than allowed, this can indicate a
malfunction in the ultrasonic transducers, electronic modules,
transducer cables or parameter settings (signal models, control
limits).
We recommend to use the default value.
1
User Warnings must be activated to become effective on the warning output.
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...