Flow Calibration
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
85
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
If the zero phase values do not meet the aforementioned criteria, the zero phase needs to
be adjusted.
To shift the zero phase cursor to the proper position, the parameter "Zerophase[rad]" in the
parameter table on the right hand needs to be adjusted.
Click the arrows to raise or lower the zero phase in steps of 0.1 rad or type in a value.
Begin by changing the zero phase value in steps of 0.5 rad.
Then click "Write to meter" and check the reaction in the signal window.
Continue until the green cursor is positioned near or exactly on the red asterisk. (The
asterisk marks the second positive zero passage.)
The proper value of the zero phase can be calculated or determined empirically.
Examples for the effect on the phase lag:
–
Zero phase + 6.28 rad (=2
):
Shifts the ZeroPhase BA (or AB) line by about one
period to the right
–
Zero phase - 3.14 rad (=-
):
Shifts about half a period to the left
After having adjusted the zero phase, check the SOS again (
).
The meter must be in the Configuration Mode with the Parameter write lock in
the UNLOCKED position.
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...