Flow Calibration
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
91
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Figure 41
"Meter calibration wizard, step 4
If new adjust factors were calculated for the meter in step 2, click "Write to meter"
(
Switch the meter to Operation Mode.
Follow the instructions on screen to verify the calibration.
Verification
Choose the test flow rate(s) between two flow rates the meter was tested at.
For example, if the errors at Qmax and 0,7*Qmax have already been found, the
correction should be tested at e.g. 0.8*Qmax. The volume measured by the
meter should be within
0.1% of the reference value. Verification for
adjustment method "Constant factor":
Verify at one test flow rate
Verification for adjustment method "Polynomial" or "Piece wise" :
Verify at two test flow rates