Flow Calibration
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
87
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
The data recorded during the calibration are stored in the meter‘s database.
Open the "Meter explorer" to see the "Measurement records" (
).
Figure 36
Meter explorer with stored calibration point records
After collecting data for all calibration points, the test results will be entered into the meter
to adjust the meter‘s output.