Maintenance report (Field setup page 8 of 8)
Create the Maintenance report.
Enter the information (Description, Technician) in the fields provided.
Specify the collection duration (e.g. 3 min.), Specify the collection duration (e.g. 3 min),
a timespan, over which live meter data is to be collected to document the meter‘s state
after the field setup. (Live data collection starts after clicking the "Start" button.)
Enter the current pressure, temperature and SOS. If the SOS is unknown, use the
Speed of sound Calculator (SOS calculator) to calculate the SOS for the gas
composition. The gas composition must be current and representative (more details
Click the "Start" button to start live data collection. Diagnosis data, measured values
and status information will be collected over the specified time span.
Once the data collection has been completed and the "Create report" button becomes
available, click on it. The Maintenance report will be created and displayed.
Print it and store a copy in the Manufacturer Data Report (MDR) shipped with the
meter.
Close the preview window.
Click the button "Close" in the Field Setup
The Field Setup is now complete.
Figure 71
Field setup page 8 of 8: Maintenance report
After creation, the Maintenance Report is automatically stored in the
MEPAFLOW600 CBM meter database. It is accessible via the "Meter explorer"
and the "Report manager". The Maintenance report can also be exported to
Excel using the direct link provided when the Maintenance report is displayed.
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...