50
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Product Description
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Table 12
Dataset type "Volume Counters"
Table 11
Dataset type "Diagnostic Values"
Unit
Description
Abbreviation
Metric
Imperial
Date/Time when data was saved
Date / Time
Sec
Sec
Flow Time
FlowTime
%
%
Average VOG
AvgVOG
m/s
ft/s
Average SOS
AvgSOS
m/s
ft/s
Profile
Profile factor
-
-
Symmetry
Symmetry
-
-
Performance per path
Performance[1..4] %
%
Turbulence per path
TurbulenceP [1..4] %
%
Average VOG per path
AvgV [1..4]
m/s
ft/s
Average SOS per path
AvgC [1..4]
m/s
ft/s
SNR per path
SNRPath [1..4]
dB
dB
AGC per path
AGCPath [1..4]
dB
dB
VOG standard deviation
VOG_stdev
m/s
ft/s
Meter Status (
)
DLOG Status
-
-
●
The dataset type "Diagnostic values" does not contain any diagnostic
information for gas velocities below the value for Vmin ("LowFlowCutOff",
Reg. #7036). The "Flow time" value shows, for the percentage of storage
cycle time the flow was above Vmin and in the flow direction specified for
the DataLog.
●
All diagnostic information is flow-weighted.
Unit
Description
Abbreviation
Metric
Imperial
Date/Time when data was saved
Date / Time
Sec
Sec
Flow Time
FlowTime
%
%
Total volume at flowing conditions (forward) Vo forward
m³
acf
Total volume at flowing conditions (reverse) Vo reverse
m³
acf
Volume at flowing conditions (forward)
Vf forward
m³
acf
Volume at flowing conditions (reverse)
Vf reverse
m³
acf
Averaged Performance
AvgPerformance
%
%
Meter Status (
)
DLOG Status
-
-
●
The dataset type "Volume Counters" does not contain values for "Averaged
performance" for gas velocities below the value for Vmin ("LowFlowCutOff",
Reg. #7036).
●
The "Flow time" value shows, the percentage of storage cycle time the flow
was above Vmin and in the flow direction specified for the DataLog.
●
The values for "Averaged performance" are flow-weighted.
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...