After the verification has been carried out successfully, go back to the first page of the
calibration wizard (
Set the pressure, temperature and meter factor values back to your customized values
(see
).
Switch the meter to Operation Mode.
LOCK the Parameter write lock (
Go to page 4 of the calibration wizard and print a "Calibration report" and a "Parameter
report" (see examples in the Appendix).
File both reports with the meter documentation.
3.5
Sealing
After completion of the calibration, carry out the following securing measures in
accordance with the sealing plan (
Close the window cover of the SPU and seal it using the securing bracket and wire lead.
Seal all transducer cover caps with at least two stickers.
Attach the conformity label to the main type plate on the SPU (