3.3
Preparation
3.3.1
Identification and visual check
The FLOWSIC600 is delivered in a pre-assembled condition in a sturdy package. When
unpacking the device, check for possible damage in transit. Pay particular attention to the
interior of the meter body, any visible transducer components and the sealing surfaces on
the flanges. Any damage must be documented and reported to the manufacturer
immediately.
For identification use the information on the main type plate (
, example see
). Cross-check the meter‘s serial number and the other information on
the main plate with the data in the Manufacturer Data Record (MDR).
Information on nominal size (DN/NPS), design temperature (TS), design pressure (PS) and
test pressure (PT) can be found on the type plate on the meter body (
, example
see
).
Figure 22
FLOWSIC600 type plates
Type plate on the meter body
Main type plate on
Signal processing unit
Содержание FLOWSIC600
Страница 94: ...94 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Flow Calibration Subject to change without notice ...
Страница 120: ...120 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Installation Subject to change without notice ...
Страница 166: ...166 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Commissioning Subject to change without notice ...
Страница 186: ...186 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Maintenance Subject to change without notice ...
Страница 200: ...200 FLOWSIC600 Technical Information 8010125 V4 0 SICK AG Troubleshooting Subject to change without notice ...