3.
To prevent interference reflections, arrange the device as close to being plane-par‐
allel to the object surface as possible.
4.
Manually place objects with bar codes one after the other into the reading range
of the device,
5.
Check the reading result with the SOPAS ET configuration software.
6.
When doing so, place objects at different positions (angles) in the reading field
and ensure that the limit values for the permitted reading angles are not
exceeded,
see "Angle alignment of the device", page 31
.
7.
Align the device so that the good read rate is between 70% and 100%.
8.
Tighten the screws on the device.
7.5
Fine adjustment and further configuration
NOTE
The other settings and the fine adjustment depend on the relevant application situa‐
tion.
User level, parameter download to the device
The user is automatically logged into the device in the “Authorized customer” level and
can change parameters which are immediately transmitted to the device (default set‐
ting).
Commissioning via Quickstart
The “Quickstart” tab offers an overview of the most important parameters and enables
quick evaluation of code content. Functions such as evaluation window, percentage
evaluation, code configuration, and adjusting tool are available via the Quickstart.
Application wizard
The application wizard (“Wand” icon) supports device configuration both as a stand-
alone device and as a master or as a slave for a master/slave combination based on
the CAN bus.
Evaluation window
The evaluation window shows the code content, the object index, the code type, the
code security, and the device number of the reading device.
Percentage evaluation
Percentage evaluation permanently assesses the quality of the reading. Bar codes are
not assessed. Here, the bar codes must not be subjected to any conveying movement.
The device performs 100 scans in each case and evaluates the reading quality. The
device continuously emits read results every 2 s via the AUX interface, together with the
read diagnostic data. A timer starts when percentage evaluation is called. If a manual
abort is not carried out, the device automatically returns to read mode after 5 min.
Adjusting Tool
The “Adjusting Tool” operating mode supports optimal placing of the center of the scan
line on the object. To do this, the device hides half of the scan line.
7
COMMISSIONING
58
O P E R A T I N G I N S T R U C T I O N S | CLV63x, CLV64x, CLV65x
8019588/129Z/2019-02-07 | SICK