- 22 -
Cycle Timer Mode
The cycle timer mode can be set only from the PC. Communications
software (sold separately) is also required.
Press the MODE key until the CY segment of the mode display bar
lights.
The flow rate setting is displayed initially in this screen.
When cycle operation has not been set, the display will be as shown
on the right, and the various setting values will not be displayed.
The cycle timer mode allows complex pump operation to be set using communications software
(sold separately) on a PC.
The sampling start setting involves only setting of the time. The sampling end time setting, however,
involves setting of three conditions, each of the time, duration (i.e. how many minutes later) and the
volume timer.
Setups comprising the above settings for up to five measurements can be registered.
In the cycle timer mode, though the time setting is the sampling start condition,
sampling is not started automatically.
Sampling is started on condition that the pump is turned ON (power is being supplied to the
pump) and that the screen is set to the "sampling standing by" screen by pressing the
START/STOP key.
Note that pump operation cannot be started simply with the pump turned ON.
Also, the cycle timer mode is designed to automatically move to the next cycle if cycle
operation is not started within two minutes of the preset time. Note also that the pump will not
operate according to the preset settings in this case.