- 16 -
Operation Modes
This pump has the following four operation modes.
(1) Manual mode operation
(See page 17.)
Pressing the START/STOP key starts and stops pump operation.
The only setup item provided is instantaneous flow rate.
(2) Down timer mode operation
(See page18.)
Pressing the START/STOP key operates the pump for the preset time.
Three setup items are provided in this mode: instantaneous flow rate, sampling start time and
sampling time (end time).
(3) Volume timer mode operation
(See page 20.)
Pressing the START/STOP key starts pump operation at the preset time. Pump operation ends
when the preset integrated flow volume value is reached.
Three setup items are provided in this mode: instantaneous flow rate, sampling start time and
sampling end integrated flow volume.
(4) Cycle timer mode operation
(See page 22.)
To use this mode, communications software (sold separately) and a PC are required.
In this mode, the pump can be operated automatically in accordance with the setup details
preset on the PC. Setups for up to five measurements can be registered in advance.
For details, refer to the communications software operation manual.
In the explanations for each of the following modes, it states that the operation display LED flashes.
However, the LED can also be set up not to flash. (See page 29.)
Содержание 090860-0341