- 27 -
Memory
Press the MODE key until the MEM screen is displayed.
In this screen, you can view past data (logs) by pressing the
START/STOP key.
"LOG 0" is displayed in the initial screen. Logs for the last ten
measurements (to LOG 9) can be viewed provided that logs are stored
in memory.
The date on this screen is the date when the pump was first operated.
* If logs are not stored in memory even after pump operation, contact
your Sibata agent.
Pressing the UP key in each log screen switches the screen cyclically and returns to the log screen
as follows: flow rate setting
→
error (displayed if occurred)
→
integrated flow volume
→
average flow
rate
→
sampling time
→
sampling start date
→
sampling start time
→
sampling end date
→
sampling end time. Pressing the DOWN key switches the screen in the reverse direction.
Which mode the pump was operated in can be checked by the mode display bar at the top of each
screen.
Pressing the MODE key at any respective position advances the screen to the next log screen.
Pressing the START/STOP key in any of the above screens returns the screen to the MEM screen.
Up to 10 logs can be checked on the mini pump body. However, up to 99 logs can be viewed on a
PC by using communications software (sold separately).
The last data displayed after pump operation stops is LOG 0.
Flow rate setting
Integrated flow volume
Average flow rate
Sampling end time
Sampling time
Sampling end date
Sampling start time
Sampling start date
Error (displayed if occurred)