8 Maintenance
8-18
8.8 Cleaning the Axial-View Observation Orifice Assembly L
WARNING
Risk of Burn Injury
There is a risk of burn injury. Do NOT touch the plasma stand interior for 3 minutes
after the plasma is extinguished.
NOTE
Cleaning the orifice assembly L helps prevent sample contamination and abnormal
electric discharge. If such phenomena are observed, clean the orifice assembly L.
The frequency of cleaning will differ depending on the samples being analyzed and
the period of use.
(1) Remove the orifice assembly L.
1) Turn the orifice assembly L with a spanner.
2) When the orifice assembly L becomes looser, turn it manually to remove it.
Tungsten tip
Apply the spanner here and turn.
Screw
Fig. 8-19 Removing the Orifice Assembly L
(2) Clean the orifice assembly L.
1) Immerse the assembly in hydrochloric acid (HCl) 2 standard solution for two minutes.
2) Remove the assembly from the hydrochloric acid and rinse it with purified water.
3) Wipe the surface (in particular, the surface of the tungsten tip) with gauze.
Содержание ICPE-9000
Страница 2: ......
Страница 6: ...iv...
Страница 13: ...xi 3 Plasma Stand Door 037 72122 00 LABEL WARN SA40 122P Japanese 037 72125 00 LABEL WARN SA40 125P English...
Страница 14: ...xii 1 3 2 Warning Labels...
Страница 16: ...xiv...
Страница 20: ...xviii...
Страница 26: ...1 Introduction 1 2...
Страница 64: ...5 Configuration and Operating Principle of the ICPE 9000 5 26...
Страница 74: ...7 Operation 7 6...
Страница 106: ...8 Maintenance 8 32...
Страница 114: ...9 Maintenance Parts and Consumable Items 9 8...