8 Maintenance
8-10
8.3.2 Attaching the Leveler
(1) Attach the drain tube.
* Insert approx. 20mm of drain tube into the bottom end of the leveler.
* Secure the drain tube with the clamp.
(2) Attach the leveler to the chamber.
* Insert the leveler into the bottom end of the chamber.
* Attach the clip for TS-15.
(3) Connect the leveler exhaust tube.
Clamp
Clip for TS-15
Leveler
Fig. 8-9 Attaching the Leveler