8 Maintenance
8-5
(1) Remove the nebulizer from the chamber.
* Hold the part connecting the chamber and the nebulizer with one hand and hold the nebulizer
with the other hand.
* Pull the nebulizer out of the chamber.
(2) Remove the carrier gas tube's clamp.
* Slide the clamp's hook up and down to remove it.
(3) Remove the carrier gas tube.
* Hold the nebulizer's carrier gas inlet with one hand and, with the other hand, pull and turn the
carrier gas tube to remove it.
(4) Remove the sample introduction tube.
* Hold the part connecting the nebulizer's sample introduction tube with one hand and, with the
other hand, pull the sample introduction tube to remove it.
8.2.3 Removing the Leveler
Clamp
Clip for TS-15
Leveler
Drain tube
Fig. 8-4 Removing the Leveler
(1) Remove the clip for TS-15.
* Hold the part connecting the chamber and the leveler with one hand and, with the other hand,
pull the clip for TS-15 to remove it.
(2) Remove the part connecting to the leveler exhaust tube.
(3) Remove the leveler.
Содержание ICPE-9000
Страница 2: ......
Страница 6: ...iv...
Страница 13: ...xi 3 Plasma Stand Door 037 72122 00 LABEL WARN SA40 122P Japanese 037 72125 00 LABEL WARN SA40 125P English...
Страница 14: ...xii 1 3 2 Warning Labels...
Страница 16: ...xiv...
Страница 20: ...xviii...
Страница 26: ...1 Introduction 1 2...
Страница 64: ...5 Configuration and Operating Principle of the ICPE 9000 5 26...
Страница 74: ...7 Operation 7 6...
Страница 106: ...8 Maintenance 8 32...
Страница 114: ...9 Maintenance Parts and Consumable Items 9 8...