5 Configuration and Operating Principle of the ICPE-9000
5-25
5.19 Optional Plasma Torch and Orifice Assembly
If axial view observations are performed using the optional plasma torch (P/N 204-70272) or high TDS
torch (P/N 204-77296), the optional orifice assembly (P/N 211-43740) is required. Replace the orifice
assembly L attached to the cooling jacket with the orifice assembly (P/N 211-43740).
Cleaning of the orifice assembly (P/N 211-43740) is the same as for cleaning of the axial-view
observation orifice assembly L in section 8.8.
Fig. 5-17 Orifice ASSY and Orifice ASSY L
NOTE
The orifice assembly L will deteriorate rapidly if the plasma is ignited using the
plasma torch (P/N 204-70272) or high TDS torch (P/N 204-77296) with the orifice
assembly L installed.
Orifice ASSY
(P/N 211-43740)
Orifice ASSY L
Turn here with a
spanner (13 mm)
Cooling jacket