5 Configuration and Operating Principle of the ICPE-9000
5-15
(1) Switching Observation Direction
The observation direction is switched using the software. When igniting the plasma, perform the
switching operation from the software's [Instrument Control] window and check that the observation
direction has been switched properly. The mark shown below is displayed in the observation
direction indicator.
Axial-view
observation:
z
Radial-view
observation:
{
(2) Switching Radial-View Observation Height
The observation height is switched using the software. The measurement position is indicated by the
observation height indicator.
The observation height for radial-view observation (i.e., the position above the RF coil) is indicated
by the observation height indicator.
(3) Changing Radial-View Observation Height
The observation height is changed using the upper and lower observation position setting knobs.
To adjust the upper observation position, loosen the locking knob with [Up] set in the software's
[Instrument Control] window, and turn the upper observation position setting knob while viewing
the observation height indicator scale. Tighten the locking knob when the position on the scale is
aligned with the desired setting.
To adjust the lower observation position, loosen the locking knob with [Down] set in the software's
[Instrument Control] window, and turn the lower observation position setting knob while viewing
the observation height indicator scale. Tighten the locking knob when the position on the scale is
aligned with the desired setting.
NOTE
Tighten the locking knob securely after adjusting the observation position. If the
locking knob is loose, the reproducibility of analytical values may be adversely
affected.
(4) Attaching Entrance Window Cover
NOTE
Remove the cooling jacket when analyzing organic solvent samples or hydrofluoric
acid samples. Otherwise, the cooling jacket may be damaged.
Attach the entrance window cover (provided) when performing radial-view observation with the
cooling jacket removed.
Содержание ICPE-9000
Страница 2: ......
Страница 6: ...iv...
Страница 13: ...xi 3 Plasma Stand Door 037 72122 00 LABEL WARN SA40 122P Japanese 037 72125 00 LABEL WARN SA40 125P English...
Страница 14: ...xii 1 3 2 Warning Labels...
Страница 16: ...xiv...
Страница 20: ...xviii...
Страница 26: ...1 Introduction 1 2...
Страница 64: ...5 Configuration and Operating Principle of the ICPE 9000 5 26...
Страница 74: ...7 Operation 7 6...
Страница 106: ...8 Maintenance 8 32...
Страница 114: ...9 Maintenance Parts and Consumable Items 9 8...