8 Maintenance
8-11
8.3.3 Attaching the Nebulizer
(1) Attach the carrier gas tube.
* Hold the nebulizer's carrier gas inlet with one hand and, with the other hand, insert approx. 10
mm of the carrier gas tube into the carrier gas inlet.
* Attach the clamp.
(2) Attach the sample introduction tube.
* Insert the sample introduction tube up to the end of the nebulizer's sample introduction tube
insertion port.
Sample introduction tube insertion port
Sample introduction tube
Clamp
Carrier gas tube
(350mm)
Carrier gas inlet
Fig. 8-10 Attaching the Nebulizer, Sample Introduction Tube, and Carrier Gas Tube
8.3.4 Attaching the Nebulizer to the Chamber
* Hold the chamber's nebulizer inlet with one hand and, with the other had, insert the nebulizer
into the chamber. Check that the straight section of the nebulizer is securely inserted in the
chamber's O-ring section.
Chamber’s
O-ring section
Straight section of nebulizer
Fig. 8-11 Attaching the Nebulizer to the Chamber
Содержание ICPE-9000
Страница 2: ......
Страница 6: ...iv...
Страница 13: ...xi 3 Plasma Stand Door 037 72122 00 LABEL WARN SA40 122P Japanese 037 72125 00 LABEL WARN SA40 125P English...
Страница 14: ...xii 1 3 2 Warning Labels...
Страница 16: ...xiv...
Страница 20: ...xviii...
Страница 26: ...1 Introduction 1 2...
Страница 64: ...5 Configuration and Operating Principle of the ICPE 9000 5 26...
Страница 74: ...7 Operation 7 6...
Страница 106: ...8 Maintenance 8 32...
Страница 114: ...9 Maintenance Parts and Consumable Items 9 8...