37
|
P a g e
USER MAINTENANCE
End of procedure
Calibration continued
7
Allow the oven to stabilize after achieving the temperature set
point using the offset display value.
Note:
The unit is stabilized when no fluctuations of
±0.2°C
or
greater
are detected.
8
Once the chamber has stabilized, compare the reference
temperature device and oven temperature display readings.
a.
If the readings are the same or the difference between the
two falls within the acceptable range of your protocol, the
display is accurately showing the chamber temperature.
The calibration procedure is now complete
.
-OR-
b.
Advance to the next step if the difference falls outside the
acceptable range of your protocol again.
Reference Device
9
Repeat steps 2 - 8 up to two more times.
•
Three calibration attempts may be required to
successfully calibrate ovens that are more than ± 2°C out
of calibration.
Reference Device
If the temperature reading difference still falls outside your protocol
after three calibration attempts, contact your distributor or
Technical
Support
for assistance.