35
|
P a g e
Begin Calibration
USER MAINTENANCE
5 ) Heat up and stabilization period
: The oven temperature must be stable at temperature in order to
perform an accurate calibration. The temperature is considered stabilized when the oven chamber has
operated at your calibration temperature for at least 30 minutes with no fluctuations
greater than
±0.2°C
.
Figure 2: Oven Chamber Heat Up and Stabilization Phases
Continued next page
Suggested Calibration Procedure
1
Once the chamber has stabilized, compare the reference
temperature device and chamber temperature display readings.
a.
If the readings are the same, or the difference between
the two falls within the acceptable range of your
protocol, the display is accurately showing the chamber
temperature.
The Temperature Calibration procedure
is now complete
.
-Or-
b.
See Step 2 if a difference falls outside the acceptable
range of your protocol.
Reference Device
2
The display requires a calibration adjustment.
•
The difference between the reference device and the
display is an
offset value
.
•
Examples of offset values:
Reference Sensor
Reading
Oven Temp.
Display
Offset
Value
152.0°C
150°C
2
149.1°C
150°C
-0.9
148.0°C
150°C
-2
•
Note the offset value for use in Step 5.
Reference Device
150°C
Start
30 Minute Minimum
Stability Period
Fluctuations
(Exaggerated)
Содержание SMO5HP-2
Страница 1: ...Installation Operation Manual SMO5HP 2 HIGH PERFORMANCE OVEN 220 240 Voltage...
Страница 8: ...8 P a g e INTRODUCTION...
Страница 12: ...12 P a g e RECEIVING YOUR UNIT...
Страница 18: ...18 P a g e INSTALLATION...
Страница 20: ...20 P a g e GRAPHIC SYMBOLS...
Страница 22: ...22 P a g e CONTROL PANEL OVERVIEW...
Страница 38: ...38 P a g e USER MAINTENANCE...