USER MAINTENANCE
Continued next page
Calibration continued
3
Jump to the Operations menu.
a.
Press and hold both the
Up
and
Down
Arrow buttons
simultaneously for approximately 5 seconds.
b.
Release the buttons when “A1” appears on the top display
line and “oPEr” appears in the mid display line.
Operations Menu
4
Advance through the Operations menu options to the Temperature
Calibration offset parameter.
a.
Push the green
Advance button
repeatedly unit “i.CA”
appears in the green mid display line and a number value in
the red top line.
5
Adjust the number value in the top display line to match the
offset value from step 2, using the arrow buttons.
6
Save the calibration offset and return to the homepage.
a.
Push the Reset button three times so the display shows the
homepage.
•
The oven will begin to heat or passively cool to reach the current set
point using the offset display value for the current temperature.
x2
x3