OPERATION
S
ET THE
H
IGH
T
EMPERATURE
L
IMIT
Note:
Test the high limit system once per year for functionality.
The high temperature limit is set by the end-user, typically at 5
°
C above the highest temperature the
oven will run at during your recipe profile or constant-temperature application.
End of Procedure
1.
Advance to the Limit High Set Point, starting from the Homepage
Push
Advance
Multiple
Times
Limit High Option
•
Push the Advance button until “Lh.S1” (Limit High Set Point) shows in the
green mid-level display line
2.
Adjust the high limit to at least 5°C above the highest temperature of your application
Adjust
Note
: If you are just checking the present high
temperature limit setting, push the Reset button to
exit the High Set Point menu and return to the
homepage without saving any changes.
3.
Save the new Limit High Setting
Push
•
The top display (red) will show “SAFE”, indicating
that the temperature limit has been saved
4.
Return to the Homepage
Push
Reset
•
Returned to Homepage