14 |
P a g e
INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
This oven is intended for use indoors, at room temperatures between
15
°
C and 40
°
C
(
59
°
F and
104
°
F)
, at no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F).
Operating the unit outside of these conditions may adversely affect its temperature range and
stability.
If this is the case, please contact your distributor to explore other oven options suited to
your laboratory or production environment.
R
EQUIRED
C
LEARANCES
These clearances are required for the oven to operate safely and meet its stated temperature
specifications
•
12 inches (305mm)
of vertical headspace clearance will suffice if the oven exhaust is vented
from the workspace through a duct or other channeling.
o
Otherwise,
24 inches (610mm)
of headspace clearance is required between the
exhaust vent and any overhead cover or partition.
•
Do not place objects on top of the oven. Exception: A power exhaust blower offered by
SHEL LAB maybe mounted on the top exhaust vent.
•
Allow at least
6 inches (152cm)
from the access port and fan vent on the back of the oven to
the nearest wall or partition. Keep the fan unobstructed at all times.
E
NVIRONMENTAL
D
ISRUPTION
S
OURCES
When selecting a location to install the unit, consider all environmental conditions that can affect its
temperature performance. For example:
•
Proximity to other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling ducts, or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
24” (620mm)
12” (305mm)
headspace with
Ducting Attached
6”
(152mm)
Door Swing
6” (152mm) Fan
12” (305mm)
6”
(152mm)
Содержание SMO5HP-2
Страница 1: ...Installation Operation Manual SMO5HP 2 HIGH PERFORMANCE OVEN 220 240 Voltage...
Страница 8: ...8 P a g e INTRODUCTION...
Страница 12: ...12 P a g e RECEIVING YOUR UNIT...
Страница 18: ...18 P a g e INSTALLATION...
Страница 20: ...20 P a g e GRAPHIC SYMBOLS...
Страница 22: ...22 P a g e CONTROL PANEL OVERVIEW...
Страница 38: ...38 P a g e USER MAINTENANCE...