www.sewhacnm.co.kr
If “UP” is displayed, please load “Test Weight” on the scale part and press
key.
Calculate Span value during 10 ~20 secs,
automatically.
※
Over than 1/10,000 resolution setting,
To guarantee the preciseness, Span calculation will be operated twice.
After calculation, span value will be displayed on
the display. Then press
key.
※This span value is not a weight value.
When “ END” is displayed and calibration is
completed.
Up
Cal-10
0.33392
end