SI 580E DIN SIZE WEIGHING CONTROLER 

www.sewhacnm.co.kr 

7. Error & Treatment

 

7-1. 

Load Cell Installation

 

Error 

Cause 

Treatment 

Remarks 

Weight Value is 

unstable 

1. Load cell broken 

2. Load cell isolation 

resistance error 

3. Weighing part 

touches other devices or 

some weight is on the 

weighing part 

4. Summing Board 

Error 

1. Measure 

input/output resistance 

of Load cell. 

2. Measure Load cell 

isolation resistance 

1. Input Resistance of 
“EXC+” and “EXC-“ is 
about 400Ω ±30 
2. Output Resistance 
of “SIG+“ and “SIG-” 
is about 350Ω ±3.5 
3. Isolate Resistance 

is more than 100MΩ 

Weight Value is 

increased regular 

rate, but not return 

to “Zero” 

1. Load cell Error 

2. Load cell connection 

Error 

1.Check Load cell 

connection 

2. Measure Load cell 

Resistance 

 

Weight Value is 

increased to under 

Zero 

Load cell Output wire 

(SIG+, SIG-) is switched 

Make wire correction 

 

“UN PASS” display 

Load cell broken or 

Indicator connection 

Error 

Load cell Check 

Load cell connection 

Check 

 

Power was “ON” when 

some weight is on the 

load cell? 

Remove weight on the 

Load cell 

 

“OL” or “UL” 

display(Over Load) 

1. Load cell broken or 

Indicator connection 

Error 

2. Loading over than 

Max. Capacity 

1. Load cell Check 

2. Load cell connection 

Check 

3. Remove over loaded 

weight 

 

 

Содержание SI 580E

Страница 1: ...Digital Weighing Indicator SI 580E Instruction Manual...

Страница 2: ...tallation 10 Page 5 Set up 11 Page 5 1 Set Up 11 Page 5 2 TEST Weight Calibration Mode 13 Page 5 3 Simulating Calibration Mode 17 Page 5 4 F FUNCTION Setting 21 Page 5 5 SET POINT Setting 5 6 Test Mod...

Страница 3: ...e in case of wrongly used Copy Rights 1 All Right and Authority for this Manual is belonged to SEWHA CNM CO LTD 2 Any kinds of copy or distribution without permission of SEWHA CNM CO LTD will be prohi...

Страница 4: ...ernal damage on item 2 Don t install under sunshine or heavy vibrated condition 3 Don t install place where high voltage or heavy electric noise condition 4 When you connect with other devices please...

Страница 5: ...60times sec Environment Operating Temperature Range 10 40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration Mode Simulation Calibration Mode Wit...

Страница 6: ...operates Lamp is ON OUT4 When OUT4 Relay operates Lamp is ON 3 2 2 Key Operation 1 Normal Mode Make Weight value as Zero F07 F08 setting 2 Calibration Mode Cancel the value or move to previous step 3...

Страница 7: ...6 Set up Mode Enter Test Mode If the printer is installed under F01 01 setting when you press this key the current valued is increased And the current weight is saved and print out altogether Refer to...

Страница 8: ...TX RS 232 option Not used Not used TX RX 4 Load cell Input EXC EXC SIG SIG SHIELD 5 External Input terminal additional 2pcs digital inputs are on the top refer to F16 F17 Standard Extended Digital Inp...

Страница 9: ...LER www sewhacnm co kr 4 INSTALLATION 4 1 External Dimension Cutting Size External Dimension mm Cutting Size mm 4 2 Installation Components SI580E Connector 5EA 3P 5P 7P x 3 Pin terminal 15EA User Man...

Страница 10: ...to the terminal correctly Do not weld near the load cells Indicators or other devices Installation The Load cells 1 You can connect Max 8pcs of same capacity Load cells at once 350 2 You have to make...

Страница 11: ...ode activated Start SET UP Mode If Lock function activated Refer F function 95 Press key four times within 2sec If P W displays input 4 characters password If Password is right SETUP Mode starts If Pa...

Страница 12: ...ode1 Analog Value key 4 times Pass word Analog difference key 4 times Pass word Key test key 4 times Pass word Test Mode2 Relay Output key 4 times Pass word Analog Output key 4 times Pass word Standar...

Страница 13: ...TARE HOLD PRINT will be reset Before processing calibration please warm up the indicator during 15 min to guarantee more preciseness Calibration Key CANCLE BACK Move to left Move to right Increase set...

Страница 14: ...sion Value After DIVI displays set Decimal point with key Whenever press Decimal point will be changed And set Division values with key Finally press key to save Max Decimal point will be 0 001 and di...

Страница 15: ...0 000 resolution setting To guarantee the preciseness DEAD weight calculation CAL00 CAL09 will be operated twice In this step if there is some force or Vibration on scale part these unstable condition...

Страница 16: ...te Span value during 10 20 secs automatically Over than 1 10 000 resolution setting To guarantee the preciseness Span calculation will be operated twice After calculation span value will be displayed...

Страница 17: ...ST weight This calibration mode uses Load cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of this Simulat...

Страница 18: ...um of each load cell s capacity and make setting with Max Capacity EX There are 4pcs of load cells and each load cell s Max capa is1 000kg Then total Max Capacitywill be 4 000kg 1 000 x 4 and you have...

Страница 19: ...omatically It takes 10sec or 20sec to get the best situation Over than 1 10 000 resolution setting To guarantee the preciseness dead weight calculation will be operated twice 5 3 5 Input Max Output Ra...

Страница 20: ...key to complete the calibration process In case of plural piece of load cells are connected the rated output will be same as single load cell s Because plural load cells are connected with parallel co...

Страница 21: ...unction and make optimal operation of SI 580E controller Starting F FUNCTION Mode Press kye 4 times Input Password Displaying SETUP press key 5etUp f01 00 ESC LEFT RIGHT INCREASE SAVE Set the function...

Страница 22: ...nce and display as Zero Auto Zero Range setting F05 00 00 99 Within the Auto Zero range weighing part is steady indicator will display current weight as Zero If the weighing part is not Steady indicat...

Страница 23: ...ing part s STEADY condition and display If you set small value indicator will take STEADY fast if you set value indicator will take STEADY slow 0 5sec per set value Display Up Date speed setting F12 1...

Страница 24: ...F21 03 Under Packer Mode 2 TARE 9 STOP F21 03 Under Packer Mode 3 TARE RESET 10 RUN STOP F21 03 Under Packer Mode 4 TARE TARE RESET 11 PRINT 5 HOLD 12 PRINT the amount 6 HOLD RESET External Input Sele...

Страница 25: ...it Mode 3 SP1 SP2 SP3 SP4 Output Setting B dry User s relay 7 Accumulating Mode1 SP1 SP2 SP3 Finish Output setting 8 Accumulating Mode2 SP1 SP2 SP3 SP4 Output setting FINISH Relay delay time setting U...

Страница 26: ...28 800bps 1 4 800bps 6 38 400bps 2 9 600bps 7 57 600bps 3 14 400bps 8 76 800bps 4 19 200bps 9 115 200bps DATA Transmition Method selection F32 0 Simplex Mode Stream Mode 1 Duplex Mode Command Mode 2...

Страница 27: ...nder F21 3 4 5 setting only Print Mode Setting Weight Unit selection F41 0 Kg 1 g 2 t Print Format selection F42 0 Continuous Print Serial No and Weight will be printed continuously 1 Single Print Dat...

Страница 28: ...relay output Set time of Average Hold F54 3 0 9 When setting Average Hold set the time unit sec Automatic Hold Reset After set time Communication Mode Setting Extended Serial Interface setting Parity...

Страница 29: ...transference Whenever Print key input DATA Transference Format selection Extended Serial port F67 0 Format 1 recommended when use external display 1 Format 2 Format 1 ID No 2 Format 3 3 CAS Format Pr...

Страница 30: ...an Change to new date DATE Y M D Check Change F91 Check Current TIME data or you can Change to new time SETUP Mode Password Key Setting Change F95 Setting the password 1 When 1 displays input 4 number...

Страница 31: ...to enter FreeFall setting mode Free fall setting mode can be set only weighing mode F21 1 F21 2 F21 6 F21 7 F21 8 Free fall value Target value If you set wrongly err 8 shows Default Free fall value i...

Страница 32: ...SP1 SP1 freefall OFF OUT 2 Current weight SP2 SP2 freefall ON Current weight SP2 SP2 freefall OFF OUT 3 Current weight SP3 SP3 freefall ON Current weight SP3 SP3 freefall OFF OUT 4 Within EMPTY range...

Страница 33: ...FF OUT 2 current weight SP2 SP2 freefall ON current weight SP2 SP2 freefall OFF OUT 3 current weight SP3 SP3 freefall ON current weight SP3 SP3 freefall OFF OUT 4 current weight SP4 SP4 freefall ON cu...

Страница 34: ...F Current weight will be reaches to SP1 SP2 800 value SP2 will be OFF Current weight will be reaches to SP1 SP1 SP3 990 SP1 will be OFF SP3 will be ON after SP1 off and delay during the t1 time T1 Fin...

Страница 35: ...eight will reaches to 950 SP1 SP4 SP1 will be OFF SP4 Finish will be ON after SP1 relay Off and delay during t1 time setting When Out3 is ON the weight value is saved T1 Finish Relay Output delay time...

Страница 36: ...eight reaches to 500 SP1 SP3 OUT3 will be OFF Current Weight reaches to 800 SP1 SP2 OUT2 will be OFF Current Weight reaches to 950 SP1 SP4 OUT1 will be OFF Within the Empty Range again SP4 will ON Whe...

Страница 37: ...FF OUT 2 current weight SP2 SP2 freefall ON current weight SP2 SP2 freefall OFF OUT 3 current weight SP3 SP3 freefall ON current weight SP3 SP3 freefall OFF OUT 4 current weight SP4 SP4 freefall ON cu...

Страница 38: ...If you set wrongly err 8 shows Default Free fall value is 0 Relay Output Relay Outut OUT 1 Current weight SP1 SP1 freefall ON Current weight SP1 SP1 freefall OFF OUT 2 SP1 Current weight SP2 SP2 free...

Страница 39: ...y err 8 shows Default Free fall value is 0 Relay Output Relay Output OUT 1 Current weight SP1 SP1 freefall ON Current weight SP1 SP1 freefall OFF OUT 2 Current weight SP2 SP2 freefall ON Current weigh...

Страница 40: ...although pressing keys or loading some force on in weighing part it may something wrong with load cell cable connector or A D board TEST MODE 1 ESC BACK Analog value Check Mode Analog Variation Value...

Страница 41: ...through Display The last digital value can be fluctuated Showing 100 000 Showing 1 000 000 Showing 10 000 000 5 6 2 Analogue Value Check Mode Under this mode you can check the variation degree of anal...

Страница 42: ...ayed on the each position 5 6 4 Relay Output Test Mode Each relay will be output by sequence Total 4pcs relay will be output one by one and circle continuous Before testing this mode please remove all...

Страница 43: ...7 Extended Serial Interface Test Mode Connect with PC or other devices through serial interface and check the transference and receipt At the normal operation display will be blinked To test this mode...

Страница 44: ...www sewhacnm co kr 6 INTERFACE 6 1 1 Serial Interface RS 422 6 1 2 Serial Interface RS 485 32 pcs of Indicators are connectable 32 pcs of Indicators are connectable 5 1 5 8 0 e 5 1 5 8 0 e 5 1 5 8 0...

Страница 45: ...sewhacnm co kr 6 1 3 Serial Interface RS 232 Serial communication interface is sensitive to electric noise Install isolated place from Power cable or other electric cables and wires and please use sh...

Страница 46: ...HT Tare is not set ST Stable GS when setting TARE US Unstable 2 Data Format2 ID Number Data Transference Refer F function 37 67 01 F18 Header1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST Sta...

Страница 47: ...1 1 1 STABLE 1 Hold Print Gross Weight TARE ZERO Header1 Header2 OL OVER LOAD NT GROSS weight ST Stable GS Net weight US Unstable 6 1 5 Command Mode Under Command Mode Indicator will recognize the re...

Страница 48: ...X ID RCWD P decimal point No 1byte DATE 6byte TIME 6byte S N 6byte TARE 7Byte 1byte current tare weight 7byte 1byte current weight 7byte weight unit 2byte ETX Ex DATE Aug 12th 2009 TIME 12 00 00 P N 1...

Страница 49: ...12th 2009 Tare Data ASCII STX ID 2Byte RTAR ETX HEX 02 30 31 52 54 41 52 03 SI 580E response STX ID RTAR P decimal point 1byte 1byte TARE vlaue 7byte ETX Ex TARE 2 000kg decimal point Set Point1 Set v...

Страница 50: ...point SP 1 2 3 4 set value ASCII STX ID 2Byte RSPA ETX HEX 02 30 31 52 53 50 41 03 SI 580E response STX ID RSP4 P decimal point 1byte SP1 set value 7byte SP2 Set value 7byte SP3Set value 7byte SP4Set...

Страница 51: ...TX ID NAK ETX Hold Reset ASCII STX ID 2Byte WHRS ETX HEX 02 30 31 57 48 52 53 03 SI 580E response normal STX ID ACK ETX error STX ID NAK ETX Print ASCII STX ID 2Byte WPRT ETX HEX 02 30 31 57 50 52 54...

Страница 52: ...ting ASCII STX ID 2Byte WTIM Time 6byte ETX Ex Time 12 00 00 SI 580E response normal STX ID ACK ETX error STX ID NAK ETX Change S N ASCII STX ID 2Byte WSNO S N 6byte ETX Ex S N is changed to 100 SI 58...

Страница 53: ...set value 7byte ETX Ex SP4 8 000kg decimal point 0 000 SI 580E response normal STX ID ACK ETX error STX ID NAK ETX Setting SP1 2 3 4 ASCII STX ID 2Byte WSPA SP1set value 7byte SP2set value 7byte SP3se...

Страница 54: ...Ex Current Weight Command RCWT ID No 01 Current Weight 1 000kg 1 P C Read Command Format STX ID NO RCWT ETX Check sum not used PC transmits to SI580E SI580E Response to PC 2 When PC requests to Indic...

Страница 55: ...000 600 ETX CHECK SUM use PC transmits to SI580E SI580E Response to PC Normal operation Incorrect operation All Read Write command must be use HEX CODE How to Calculate Check sum Sum the value from ST...

Страница 56: ...ut terminal if the high voltage power will be connected with output terminal it may cause damage or relay or main board of indicator 6 2 Relay Output 4pcs Control Relay output is installed in Output T...

Страница 57: ...log Output Setting mode Step3 Adjust the displaying value of indicator with keys by 9 999 to make Digital multi meter s value as minimum ex 4mA and save When the SI580E indicate about 5 200 the Digita...

Страница 58: ...GITAL MULTI METER How to adjust analog output value Step1 Connect Digital multi meter to the Indicator A out terminal Step2 Enter F81 Minimum Analog Output Setting mode Step3 Adjust the displaying val...

Страница 59: ...is already programmed and fixed with SE7200 7300 model 6 5 1 Printing Format Using the RS 485 or 422 interface please use convertor and converts to RS 232 and connect with Serial printer If you use RS...

Страница 60: ...out 350 3 5 3 Isolate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Measure L...

Страница 61: ...connection error 3 Test weight is not loaded Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small Eer...

Страница 62: ...connect the indicator s A D with the other indicator 4 Check the power and connection of terminal UNpa55 1 Power is ON when some materials are on weighing part Under Normal Mode if there are more than...

Страница 63: ...ir without Sewhacnm s permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non Authorized Distributo...