5-2-4. Measure the “DEAD” Weight of Weighing Scale.
When “DEAD” displays, press
key, then indicator will calculate Dead weight of scale part
automatically.
Indicator will search “DEAE weight” during 10~20
secs automatically to find the best condition.
※
Over than 1/10,000 resolution setting,
To guarantee the preciseness, DEAD weight calculation (CAL00~CAL09) will be operated twice.
In this step, if there is some force or Vibration on scale part, these unstable conditions
will be continued, “ErrorA” will be displayed, and “DEAD value” will not be calculated.
Under this condition, please remove the cause of force or vibration and process it again.
5-2-5. Input Test Weight value and Calculate SPAN value.
If “SPAN” is displayed, input “Test Weight” capacity and press
key.
dead
Cal-00
5paN
1000.0