8
Figure 2 – Flow Diagram
Figure 3 – Analyser Cell Assembly
The surfaces of both electrodes in the sensing cell assembly are kept clean by the continuous
action of PVC spheres agitated by a motor-driven rotating striker assembly. This constant cleaning
minimizes signal drift and recalibration, and provides an accurate residual measurement. The
sensing cell assembly, including stepper motor, motor cable with quick-disconnect for connection
to electronics enclosure, cathode, anode, PVC spheres, and agitator assembly can be removed
from the wet-end of the analyser as a single assembly, for easy maintenance. A 100-ohm RTD
compensates for temperature variations.