14
2.3.2
Sample Line Taps
Sample line taps into larger pipes should be installed to minimise the chances of air
bubbles or sediment entering the analyser. A tap should project into the centre of the line.
Refer to Figure 9.
2.3.3 Sample
Flow
A typical installation may require a sample pump if the pressure is low and/or a Y-strainer
to remove any particulates. There may also be a need for a pressure reducing valve if
the sample pressure is too high or the sample pump needs to be regulated. Refer to
Figure 10.
2.4 Electrical
WARNING: FAILURE TO DISCONNECT POWER TO THE Cl1000 RESIDUAL ANALYSER
PRIOR TO ACCESSING THE INTERIOR OF THE ELECTRONICS ENCLOSURE MAY
RESULT IN SERIOUS PERSONAL HAZARD FROM EXPOSURE TO LETHAL VOLTAGES.
CAUTION: ALL WIRING MUST COMPLY WITH APPLICABLE LOCAL AND NATIONAL
ELECTRICAL CODES.
2.4.1
Wall Panel Mounted Conduit Connections - Refer to Figure 11
a.
Remove four (4) screws securing the terminal cover located below the computer
screen.
WARNING: POWER FOR THE UNIT SHOULD BE SUPPLIED FROM FUSED
DISCONNECT
(
2 Amp anti surge MCB recommended
)
. THIS DISCONNECT
SHOULD BE LOCATED IN CLOSE PROXIMITY TO THE UNIT AND SHOULD
BE LABELED AS THE DISCONNECT FOR THE UNIT. WIRING SHOULD
FOLLOW ALL APPLICABLE LOCAL ELECTRICAL CODES AND
REGULATIONS.