MultiExact Gas Analyser
9 – Fault finding
Message
Measurement
screen icon
Recommended actions
High flow alarm
Configurable
Flow risen above the set level for the flow alarm:
Check that there is not excessive gas flow into the
transducer (see Section 2.4).
Check that the level has been set correctly and that
calibration has been performed (see Section 5.12).
Check sample is clean and without contaminants.
Low flow alarm 1 /
Low flow alarm 2
Configurable
Flow has fallen below the set level for the flow
alarm:
Check that there is sufficient gas flow into the
transducer (see Section 2.4).
Check that the level has been set correctly and that
calibration has been performed (see Section 5.12).
Check the valve block cable is fitted correctly
(see Section 5.1.4).
mA fault
Ensure that the electrical cabling connected to the
analyser is not open circuit.
Turn the analyser off, and then turn it on again. If
the fault persists, contact Servomex or your local
Servomex agent for assistance.
mA not detected
Contact Servomex or your local Servomex agent
for assistance.
mA overridden
Raised during checking of the mA output
(see Section 7.4).
mA reset
Contact Servomex or your local Servomex agent
for assistance.
P calib fault
Calibrate the pressure compensation transducer
again.
P high fault
Adjust pressure. Contact Servomex or your local
Servomex agent for assistance.
P H/W fault
Contact Servomex or your local Servomex agent
for assistance.
P low fault
Adjust pressure. Contact Servomex or your local
Servomex agent for assistance.
Power config fault
Contact Servomex or your local Servomex agent
for assistance.