MultiExact Gas Analyser
Appendix
Register
Name
Comments
1137
Transducer calibration
mode
Calibration mode is active.
1138
Auto validation/calibration.
Sequence is in progress.
1139
mA service in progress
Calibration or override in progress.
1140
Volt service in progress
Calibration or override in progress.
1141
Pressure service in progress Calibration in progress.
1142
Relay service in progress
Override in progress.
1143
Flow service in progress
Calibration in progress.
1144
No Profibus module
Hardware fault or card missing
1145
High flow alarm
Flow is above the high flow alarm level
1146
Flow temperature fault
Flow hardware has a temperature fault
1147
Flow cal high diff
Zero and normal flow calibration points
are too far apart
1148
Flow cal low diff
Zero and normal flow calibration points
are too close together
1149
Transducer error
Transducer error
1150
Temperature fault
Temperature fault
05410001A / Revision 6
111