Appendix
MultiExact Gas Analyser
Register
Name
Comments
138
Start year
0 – 99
139
Start month
1 – 12.
140
Start date
1 – 31.
141
Start hour
0 – 23.
142
Start minute
0 – 59.
143
Repeat hour
Integer.
Pressure
144
Compensating
0 = no, 1 = yes.
Alarm set up
145
Alarm 1 mode
0 = none, 1 = low, 2 = high.
146
Alarm 1 latching
0 = no, 1 = yes.
147 – 148
Alarm 1 level
Float.
149 – 150
Alarm 1 hysteresis
Float.
151
Alarm 2 mode
0 = none, 1 = low, 2 = high.
152
Alarm 2 latching
0 = no, 1 = yes.
153 – 154
Alarm 2 level
Float.
155 – 156
Alarm 2 hysteresis
Float.
Flow Alarm
157
Low flow 1 level
A status will be raised if flow falls below this
level.
158
Low flow 2 level
A status will be raised if flow falls below this
level.
159
Low flow 1 status type The type of status reported for low flow
alarm 1
0 = None, 1 = Fault,
2 = Maintenance required.
160
Low flow 2 status type The type of status reported for low flow
alarm 2
0 = None, 1 = Fault,
2 = Maintenance required.
161
High flow status type
The type of status reported for high flow
alarm
0 = None, 1 = Fault,
2 = Maintenance required.
114
05410001A / Revision 6